Auditdata SWAP service

Auditdata SWAP service

Instructions for the proper transducer exchange in place

Introduction

This document explains how to ensure that the new transducer was recognized by Primus fitting unit.

Instructions


  1. Connect the Fitting Unit to the PC using the USB cable.
  2. Start Primus.
  3. Wait until the Fitting Unit is initialized and connected. The current connection status is displayed on the right side of the status bar at the bottom of the screen.
    \
  4. Disconnect the old transducer.
  5. Connect the new transducer to the Fitting Unit.
  6. In Primus, go to Help > License Information > Connected Devices.
  7. Review the name and serial number of the new transducer.
    If the data corresponds to the new transducer, you can start using the system normally.
    If the data corresponds to the old transducer, proceed with the next steps. 
  8. In Primus, go to Tools > Refresh Transducer Data.
  9. Select the needed transducer and click Refresh Transducer Data.
  10. Wait until this process is finished and close this window.
  11. You can close the window and start using the system normally.

    • Related Articles

    • Video - Adding speech material to Primus and distribute via cloud

      This short video will demonstrate how to upload the Speech Material into Auditdata Cloud for sharing throughout an organisation. ​
    • How do I share Test Definitions from Auditdata Cloud?

      You can share Test Definitions in a very similar way to sharing Primus Settings Profile described in the previous section.  To do this:  ​ Please ensure that you have customized your Test Definitions in Primus under Tools > Test Definitions.  Go ...
    • Technical Instructions for Use & User Manual - Primus

      This document aims to provide instructions on setting up Primus and installing and configuring the Primus software. It also includes essential information on safety measures, maintenance, and calibration. The Fitting Unit is a general term combining ...
    • Audiometry measurement has limited range on graph in Primus software

      After transducer changing it might happen that a limitation of the measurement range is present on graph in Primus software. The described issue relates to misbehavior of older Primus software versions and new transducer plugs. Primus software ...
    • Primus downloads and release notes

      We're thrilled to announce some transformative changes to our beloved Primus Software! Driven by our commitment to continuously enhance your user experience, we've made significant updates that we're confident you'll find remarkable. Rebranding: ...