Error P0077 when starting Symphony 1.3

Error P0077 when starting Symphony 1.3

1. Uninstall Symphony software from Control Panel of Windows.
2. Open this folder:
C:\Users\Public\Documents\Otovation
Once inside this folder, find and delete the Symphony folder
3. Reboot PC.
4. Make a right click on the Symphony installation file and select "Run as administrator".

5. Finish installation of Symphony.
    • Related Articles

    • Patient Communications Not Working

      1. Patient communications is not supported for USB connections and for Amplitude T-series audiometers. 2. If you are using internal Bluetooth, then the type of internal Bluetooth may not be compatible with Symphony. Refer to the Quick Start Guide for ...
    • Unable to load DLL 'SqlServerSpatial110.dll': The specified module could not be found

      After Symphony 3 installation and it can't be opened. The error "Unable to load DLL The specified module could not be found" appears on start and also is visible in Symphony.log Solution: 1. Visit official Microsoft website to access Microsoft® SQL ...
    • Lost connection

      The audiometer can lose its connection from Symphony when the Bluetooth signal is either lost or weak. Bluetooth communications are limited to a distance of 30 ft. (~ 10 meters), so the audiometer must be within this range in order to connect to ...
    • No Bluetooth Radio was found

      If you get this message, Symphony cannot detect the Bluetooth radio on your computer.  1. Open the Task Manager:          a. Press the CTRL, SHIFT, and ESC keys to open the Task Manager.          b. Click on the Services tab.          c. Right-click ...
    • OTOPod Service is not running

      If you get this message, the OTOPod Service failed to start with Windows. To re-start the service, perform the following: 1. Open the Task Manager:          a. Press the CTRL, SHIFT, and ESC keys to open the Task Manager.          b. Click on the ...