Failed USB Connection

Failed USB Connection

Press the power button on the audiometer again, wait for the slow blinking LED light, and then click on the Connect button a second time. If the connection fails a second time, check the following:

1. Open the Device Manager by clicking on the Device Manager button next to the Connect button:

         a. Verify that the serial number matches the device.

         b. Verify that the connection method is “USB.”

         c. Verify that the audiometer is specified as the Default audiometer.

2. Ensure that the USB cable is properly attached to the OTOPod and a USB port on the computer.

3. Try unplugging and re-inserting the USB cable into the computer. There should be some kind of audible or visual response to indicate that Windows recognizes the USB device.

4. If the audiometer LED light blinks rapidly (faster than once per second) for an extended period of time, the batteries may be low. Attach the audiometer to external power using the power supply and re-connect.


    • Related Articles

    • Lost connection

      The audiometer can lose its connection from Symphony when the Bluetooth signal is either lost or weak. Bluetooth communications are limited to a distance of 30 ft. (~ 10 meters), so the audiometer must be within this range in order to connect to ...
    • Failed Bluetooth Connection

      In some cases, the audiometer may fail to connect due to Bluetooth and/or device readiness. In this case, press the power button on the audiometer again, wait for the slow blinking LED light, and then click on the Connect button a second time. If the ...
    • Patient Communications Not Working

      1. Patient communications is not supported for USB connections and for Amplitude T-series audiometers. 2. If you are using internal Bluetooth, then the type of internal Bluetooth may not be compatible with Symphony. Refer to the Quick Start Guide for ...
    • Error P0077 when starting Symphony 1.3

      1. Uninstall Symphony software from Control Panel of Windows. 2. Open this folder: C:\Users\Public\Documents\Otovation Once inside this folder, find and delete the Symphony folder 3. Reboot PC. 4. Make a right click on the Symphony installation file ...
    • OTOPod Service is not running

      If you get this message, the OTOPod Service failed to start with Windows. To re-start the service, perform the following: 1. Open the Task Manager:          a. Press the CTRL, SHIFT, and ESC keys to open the Task Manager.          b. Click on the ...