Both Primus and your OMS system should work on the same workstation.
1. Open Primus and go to Tools - Settings - Common section.
Enable GDT Interface and put in configuration details (please, see the screenshot for the example):
2. Open your OMS system and configure GDT on it as well.
Please note, that you need to use the same file names however they need to be opposite (receiver becomes sender and sender should be receiver).
If everything is set properly you must have the following workflow:
1. You select a patient in your management system and generate GDT request for Primus.
2. A new file
receiver.gdt should be created in the specified folder.
3. Once Primus is open it automatically reads the file and it disappears from the folder.
4. You perform some measurements in Primus and save the results.
5. A new file
sender.gdt should be created and automatically read by your management system.
6. The file disappears accordingly.
GDT works with Pure Tone and Speech only.