Primus does not recognize transducer - "Sound Booth Cable Connection" message

Primus does not recognize transducer - "Sound Booth Cable Connection" message

IMPORTANT
Please note that some transducers will only work with a Fitting unit that run with firmware version equal: 3.2.0.0.0247 or higher and Primus version 4.0
Check firmware procedure:
1. Please start the Audiometer software with the Fitting Unit connected to the PC
2. Go to the menu: “Tool” at the top
3. Select: “Update Firmware” in the menu
4. Under Audiometer hardware check that Current firmware version is equal or higher than: 3.2.0.0.0247 Installation procedure:
5. If this is not the case, please install the newest Audiometer software Primus version 4.0 that will prompt you to install the latest firmware
NOTE
1. If you get the following dialog in Audiometer software the: “Sound Booth Cable Connection”. This means you are running with an older firmware version that the transducer does not support, therefore update the firmware using the latest Audiometer software as above.

    • Related Articles

    • What is the process for switching transducer cables in Primus?

      Connect the transducer with the original cable – ONLY this transducer – to the Primus unit. In Primus, choose ‘Tools’ → ‘Copy Transducer Data’, and follow the on-screen instructions: Press Next Press Next When ‘Reading plug information..…’ is ...
    • Primus Fitting Unit (PFU) is displayed as Not connected

      The status bar in Primus when device is Not connected: Checklist: Reconnect a USB cable and power supply Try another USB port on the computer Try another USB cable If a connection is via hub/switch, connect directly to the computer Check that the ...
    • Message is displayed "Another instance of application is already running" when opening Primus

      The message appears when Primus has been closed recently and the PC system didn't reflect that (due to low memory). To force quit Primus in such case, please open Task Manager and go to Details tab. Select Audiology.FittingSoftware.Standalone... and ...
    • Hardware error displayed in Primus

      1. Run hardware diagnostic test Enable Hardware Diagnostic Test in Tools > Settings > General > Common   Start the test under Tools > Hardware Diagnostic Test Select the device (Measurement or Hearing Instrument), all needed parameters, and ...